A flashing or steady green LED indicates the unit is charging or in "floating operation" mode.
operates via an integrated rechargeable battery system, making it suitable for mobile use in ambulances or outdoors.
Disinfect paddle electrodes with a commercial disinfectant like Gigasept FF. primedic defi b service manual work
Inspect the paddle cable insulation for fraying or damage.
Ensuring the device reaches max energy in under 15 seconds. A flashing or steady green LED indicates the
Ensure paddles are clean and free of residual electrode gel. Dried gel can cause sparking or poor contact during a shock. 3. Cleaning and Disinfection
Connect the power cord while the main switch is in the "OFF" position. Inspect the paddle cable insulation for fraying or damage
Verifying output at 10J, 50J, 100J, 200J, 300J, and 360J (± 15%).